1.0 OBJECTIVE :
- To provide a procedure for Operation of Micrometer Screw Gauge.
2.0 SCOPE :
- This procedure is applicable to Micrometer Screw Gauge.
3.0 RESPONSIBILITY :
- Quality control Chemist and above
4.0 ACCOUNTABILITY:
- Quality Control Head
5.0 PROCEDURE :
- Micrometer Screw Gauge is made up of a main scale and a thimble scale and it is used for measuring the thickness of packaging materials like cartons, aluminum foil etc.
- It can measure maximum thickness of 25.0 mm.
- The thimble has a total of 50 markings representing 0.50 mm.
- The smallest marking is 1 mm on the main scale (sleeve) and 0.01 mm on the thimble scale (thimble).
- Least count of Micrometer Screw Gauge is 0.01 mm.
5.1 OPERATION
- Set the 0 by gently moving the thimble.
- In case it is not zero, then instrument has zero error which is be adjusted.
- Put the object between the fix rod and movable rod and fix the object between two rods gently moving the screw.
- Spin the movable rod clockwise until the spindle just meets the object.
- Set the lock nut while the micrometer is still on the object to fix the position.
- Note down the reading of scale as below:
- The first part is contributed by the main scale on the sleeve.
- To obtain the first part of the measurement: Look at the image image (b) you will see a number 5 to the immediate left of the thimble.
- This means 5.0 mm. Notice that there is an extra line below the datum line, this represents an additional 0.5 mm. So the first part of the measurement is 5.0 + 0.5 =5.5.
- The second part is contributed by the rotating vernier scale on the thimble.
- To obtain the second part of the measurement: Look at the image image (b), the number 28 on the rotating vernier scale coincides with the datum line on the sleeve.
- Hence, 0.28 mm is the second part of the measurement.
- You just have to add the first part and second part of the measurement to obtain the micrometer reading,
- i.e., 5.5 + 0.28 = 5.78.
- Remove the object from the micrometer.
- Reset the lock nut to the unlock position.
- Keep the instrument in box after use.
- Calibration of Micrometer Screw Gauge is perform by approved outside agency on yearly basis.
6.0 ABBREVIATIONS:
Abbreviation | Expanded form |
mm | Millimeter |
7.0 ANNEXURES:
- Nil
8.0 SOP REFERENCES
- Manual of Micrometer Screw Gauge
END OF THE SOP