Tap Density Tester SOP

Tap Density Tester Calibration Procedure | Tap Density Tester Electrolab Operation Procedure | Bulk Density Tester |

  • Tap Density Tester SOP covers below points :
    • Operation of Tap Density Tester
    • Precautions
    • Calibration of TAP Density Tester
    • Format For Calibration Of Tap Density Tester
1.0 OBJECTIVE :
  • To lay down a procedure for the operation and calibration of Tap Density Tester.
2.0 SCOPE :
  • This SOP is applicable for the operation and calibration of Tap Density Tester used for measuring tapped density of samples.
3.0 RESPONSIBILITY :
  • The QC Staff shall be responsible for:
    • Following the procedure for operation of Tap Density Tester.
    • Following the procedure for calibration of Tap Density Tester.
  • The Manager-QC shall be responsible for:
    • Ensuring compliance of the SOP
4.0 ACCOUNTABILITY:
  • Quality Control Head
5.0 PROCEDURE :
5.1 Operation of Tap Density Tester:
  • Place the cylinder holder on the anvil of the desired method on instrument.
  • Switch on the power and then the instrument by using the switch provided on the rear side of it.
  • Using digit scroll key set the mode of operation as USP or USER.
  • Set the method for the test as USP I or USP II using METHOD key. 
  • Press the SET key and set the tap count I (500 taps as per USP) by using the UP/DOWN and DIGIT scroll key. Then press the enter key to go to the next parameter i.e. tap count II and tap count III.
  • Set the tap count II (750 taps as per USP) and tap count III (1250 as per USP) using UP/DOWN and DIGIT scroll keys. Then press the enter key.
  • Press the enter key and ensure that the time is displayed on the screen.
  • Again press the enter key and ensure that the display shows date on the screen. Set the date if required using the UP/DOWN digit scroll key.
  • Press the enter key and wait till the display shows ‘PRINT LAST RESULT? Y/N’. Set it to Y or N as required.
  • Press the set key to exit from the set menu.
  • Press the start key and enter the weight of sample.
  • Fill weighed sample in cylinder and lock it properly. Then press the enter key and note down the volume. Enter it (Vo) using UP/DOWN and DIGIT keys.
  • Press start key to run the test.
  • After the programmed tap count is over, measure the volume (Va) to its nearest graduated unit. Then press the enter key.
  • Press the start key to continue Tap count II and measure the tapped volume (Vb).
  • If the difference between the two volumes is less than 2 % then Vb is the final tapped volume, otherwise continue the test until the difference between succeeding measurements is less than 2%.
  • Press the stop key and note down the tapped density displayed on the screen. Similarly calculate the initial density, Hausner ratio, compressibility index etc.
  • Take the print of completed test reports and press the set key from the front panel to initialize the instrument for a new test.
  • Make entries in log book.

  • Precautions:
    • Keep the shaft and the tip of the cylinder holder clean.
    • Place the cylinder in the holder and make sure that it is properly locked.
5.2 Calibration of TAP Density Tester :
  • Operate the instrument as per operating procedure.
  • Select method 1 and set drops per minute as 300. Operate instrument and count the number of drops per minute using stopwatch. Note down the counts in calibration protocol.
  • Select method 2 and set drops per minute as 250. Operate instrument and count the number of drops per minute using stopwatch.
  • Note down the counts in calibration protocol

  • Acceptance Criteria:
    • Drop per minutes should be 300 ± 2%.
6.0 Abbreviations:
AbbreviationExpanded form
SOPStandard Operating Procedure
No.Number
Q.C.Quality Control
QAQuality Assurance
DeptDepartment
No.Number
%Percentage

7.0 Annexures:
Annex. No.Title
01Format For Calibration Of Tap Density Tester

8.0 SOP References:
  • In-House

END OF THE SOP


ANNEXURES :

Annex. No. 01 Format For Calibration Of Tap Density Tester