HPLC Calibration SOP | Calibration of HPLC by Caffeine | HPLC Calibration Parameters | HPLC Calibration Procedure |
- HPLC Calibration SOP covers below points:
- Precautions
- HPLC Calibration parameters
- Performance of pump for constant flow rate (pump calibration)
- Wavelength accuracy test for UV detector
- Linearity of detector response
- Calibration of injector & loop
- Precision and Linearity of Injector and Loop
- Auto Sampler Injector Calibration
- Calibration for gradient program test
- Column oven temperature
- Carry over test
- HPLC Calibration format
1.0 OBJECTIVE :
- To define the Calibration procedure for High Performance Liquid Chromatography.
2.0 SCOPE :
- This procedure is applicable to HPLC.
3.0 RESPONSIBILITY :
- Quality Control Officer and above
4.0 ACCOUNTABILITY:
- Quality Control Head
5.0 PROCEDURE :
5.1 PRECAUTIONS
- Ensure that the plug is connected to the mains.
- Check and ensure that the instrument is clean and calibration is valid.
- Ensure that the pump, column, detector & processors are properly connected.
- Set mobile phase and waste solution bottle in place.
- Ensure that insert the end of solvent tube into solvent bottle.
- Ensure that HPLC is connected with UPS and have battery backup.
- Before you start the analysis of a sample. Ensure that ,
- Suitable column is installed,
- System is purged and equilibrated,
- Solvent bottles are filled by the mobile phase or by solvent,
- Confirm the wavelength, flow rate , column temperature etc.
5.2 CALIBRATION
- Perform the calibration of instrument as per procedure given below.
- Frequency for calibration of HPLC system: Half Yearly.
- PERFORMANCE OF PUMP FOR CONSTANT FLOW RATE
- Start the pump with ON/OFF key.
- Purge all the port with degassed water for 10 minutes.
- Connect the Union Instead of Column.
- Select Pump – A from instrument for calibration.
- Set the flow 0.25 ml /min. and start the pump to stabilize the system.
- When the system gets stabilized, Weigh the empty 20 ml volumetric flask and put at the outlet of Detector end. Start the stopwatch immediately after starting the run.
- After 5 minutes, take out the volumetric flask and weigh the volumetric flask containing water, measure the time by using calibrated stop watch.
- Calculate the flow deliver by pump per minute using the density of water 0.99602 gm/ml at 25°C by using following formula.
- Actual flow = Weight of water in gm / Time X 0.99602
- Where 0.99602 is density of water at 25°C.
- Repeat twice the same steps
- Repeat above steps by changing flow rate at 0.5 ml/min, 1.0 ml/min, 1.5 ml/min., 2.0 ml/min. and actual flow delivered by pump A is determined.
- Repeat above steps for Calibration of Pump-B, Pump-C and Pump-D.
- Confirm and record the results.
- Acceptance criteria: Mean of three replicate should be as follows;
Flow rate | Acceptance criteria |
0.25 ml / minute | 0.25 ml ± 0.1 ml |
0.5 ml / minute | 0.5 ml ± 0.1 ml |
1.0 ml / minute | 1.0 ml ± 0.1 ml |
1.5 ml / minute | 1.5 ml ± 0.1 ml |
2.0 ml / minute | 2.0 ml ± 0.1 ml |
- WAVELENGTH ACCURACY TEST FOR UV DETECTOR:
- CHROMATOGRAPHIC CONDITIONS:
- The chromatographic conditions are as follows;
- Column : C18(250 mm × 4.6 mm)5 µm
- Column temperature : 250C
- Flow rate : 1.0 ml/ min
- Mobile phase : Methanol (70) : HPLC water (30)
- Injection Volume : 20 µl
- Prepare the 20 ppm of caffeine in HPLC water as follow;
- Caffeine Stock Solution: Weigh accurately about 50 mg of Caffeine In 100 ml volumetric flask, add 60 ml of HPLC water. Sonicate for 15 min. and diluted to volume with HPLC water.
- Dilute 2 ml of Caffeine stock solution to 50 ml with HPLC water.
- Inject the preparation at 269, 270, 271, 272, 273,274, 275, 276 and 277 nm.
- Same time check that the maximum area should be detected at the wavelength 272 nm ± 2.0 nm.
- Confirm and record the results.
- LINEARITY OF DETECTOR RESPONSE
- Chromatographic conditions: The chromatographic conditions are as follows;
- Column : C18(250 mm × 4.6 mm)5 µm
- Column temperature : 250C
- Flow rate : 1.0 ml/min.
- Mobile phase : Methanol (70) : HPLC water (30)
- Wavelength : 272 nm
- Injection volume : 20 µl
- Then prepare the level solution of 20, 30, 40, 50 and 60 ppm with the help of the dilution of Caffeine stock solution.
- 20 ppm solution: Dilute 2.0 ml of the Caffeine stock solution to 50 ml with HPLC water.
- 30 ppm solution: Dilute 3.0 ml of the Caffeine stock solution to 50 ml with HPLC water.
- 40 ppm solution: Dilute 4.0 ml of the Caffeine stock solution to 50 ml with HPLC water.
- 50 ppm solution: Dilute 5.0 ml of the Caffeine stock solution to 50 ml with HPLC water.
- 60 ppm solution: Dilute 6.0 ml of the Caffeine stock solution to 50 ml with HPLC water.
- Inject the each solution three times; calculate the mean area RSD%. (NMT 2.0%)
- Plot the graph of concentration against area for the linearity of detector response.
- Calculate the correlation coefficient and correlation coefficient should not be less than 0.998.
- Remark the performance of detector in terms of linearity is satisfactory or not & note the results.
- CALIBRATION OF INJECTOR & LOOP
- Precision and Linearity of Injector and Loop
- Chromatographic conditions: The chromatographic conditions are as follows;
- Column : C18(250 mm × 4.6 mm)5 µm
- Column temperature : 250C
- Flow rate : 1.0 ml/min.
- Mobile phase : Methanol (70) : HPLC water (30)
- Wavelength : 272 nm
- Injection volume : 5 µl ; 10 µl; 20 µl; 50 µl and 100 µl
- Prepare the 20 ppm of caffeine in HPLC water by diluting 2 ml of stock solution to 50 ml with HPLC water and inject the triplicates of 5 µl, 10 µl, 20 µl, 50 µl and 100 µl of sample preparation.
- Plot the graph of Injection volume against area.
- Calculate correlation coefficient should not be less than 0.998.
- Remark :The performance of Injector & Loop in terms of Linearity and Precision is satisfactory or not & note the results.
- Chromatographic conditions: The chromatographic conditions are as follows;
- Auto Sampler Injector Calibration
- Chromatographic conditions: The chromatographic conditions are as follows;
- Column : Union
- Flow : 1.0 ml / min
- Wavelength : 254 nm
- Mobile phase : Water
- Injection volume : 5 ml/10 ml/20 ml/50 ml/100 ml
- Run time : 2 minutes
- Fill HPLC vial with HPLC water .
- Weigh accurately filled vial & note down the weight of filled vial (A).
- Inject 10 injections of equal volume about 5 ml of HPLC water from the same vial.
- Reweigh the same vial & note down the weight (B).
- The difference in the weighing gives the weight of water injected. Maintain temperature at 25°C during whole procedure.
- Determine the volume of water injected by assuming that the weight of 1 ml of purified water at 25°C when weighed in air of density 0.0012 gm per ml is 0.99602.
- Volume of water per Inj. = A – B x 0.99602 x 1000 / 10
- Repeat the same procedure by taking injection volume 10 ml, 20 ml, 50 ml & 100 ml instead of 5 ml.
- Acceptance criteria: Volume in ml of individual injection should be ± 2% of defined volume.
- Chromatographic conditions: The chromatographic conditions are as follows;
- CALIBRATION FOR GRADIENT PROGRAM TEST
- High pressure gradient elute delivery of 2 mobile phases (HPLC water and 10 ppm caffeine solution) is performed, the absorbance produced by various set concentrations is measured and the measurements are used to calculate the actual concentrations.
- Purge all the port with HPLC water.
- Replace the column with Union.
- Equilibrate system up to minimum 1 hour.
- Before gradient program test is carried out, Parameters set are as below,
- Pump A – C Receiver : 500 ml HPLC water.
- Pump B – D Receiver : 500 ml 10 ppm caffeine
- Pumping mode : Gradient system ( A, B, C, D )
- Flow Rate : 1 ml/min
- UV Detector : 272 nm
- Temperature : 25°C
- Stop Time : 35 minutes.
- GRADIENT PROGRAME TEST TIME
- Calculate the observed mobile phase composition Cx (in % B and in % D ) from Hx values by the following formula
- Cx (%) = Hx x 100 / H100
- Where,
- X : For programmed step height (10%,50%,90%,100%)
- Hx : Height of respective plateaus in mm
- Cx : The observed mobile phase composition
- H100 : Height for 100% peak.
- Cx (%) = Hx x 100 / H100
- Acceptance criteria : The observed mobile phase composition (Cx) shall be within ± 2.0 of the set composition.
- Remark the performance of gradient composition as satisfactory or not & note the results.
- COLUMN OVEN TEMPERATURE
- Place a Calibrated digital temperature thermometer probe inside the column heater.
- Ensure that the thermometer probe does not contact any inside surface of oven.
- Close the column compartment properly.
- Set column compartment temperature 30°C .
- Take sufficient time to the oven temperature to reach and stabilize at set temperature.
- Record the result.
- Repeat the procedure for 40°C & 60°C and record the results.
- CARRY OVER TEST
- Chromatographic conditions: The chromatographic conditions are as follows;
- Column : C18(250 mm × 4.6 mm) 5 µm
- Column temperature : 25°C
- Flow rate : 1.0 ml/min.
- Mobile phase : Methanol (70) : water (30)
- Wavelength : 272 nm
- Injection volume : 20 µl
- Sample preparation: 100 ppm solution : Dilute 10 ml of Caffeine Stock solution to 50 ml with HPLC water.
- Calculate carryover by the formula;
- Carryover (%) = Carryover Area in blank x 100 / Test Area
- Acceptance criteria: Not more than 0.01%.
- Chromatographic conditions: The chromatographic conditions are as follows;
6.0 ABBREVIATIONS:
Abbreviation | Expanded form |
ml | Milliliter |
min | Minute |
0C | Degree Celsius |
nm | Nanometer |
% | Percentage |
Abs | Absorbance |
UPS | Uninterruptible Power Supply |
ppm | Parts Per Million |
gm | Gram |
mm | Millimeter |
µl | Microliter |
µm | Micrometer |
µ | Micron |
7.0 ANNEXURES:
Annex. No. | Title |
01 | Calibration Of High Performance Liquid Chromatography |
8.0 SOP REFERENCES
- Manual of High Performance Liquid Chromatography
END OF THE SOP
For HPLC Basics- Principle and Scope click link: HPLC Basics- Principle and Scope
ANNEXURES :
Annex. No. 01 Calibration Of High Performance Liquid Chromatography