1.0 OBJECTIVE :
- To provide a procedure for Operation of Digital Micrometer.
2.0 SCOPE :
- This procedure is applicable to Mitutoyo make Digital Micrometer.
3.0 RESPONSIBILITY :
- Quality Control Chemist and above
4.0 ACCOUNTABILITY:
- Quality Control Head
5.0 PROCEDURE :
- The Digital Micrometer is a precision instrument that is used to measure thickness of packaging materials like PVC film, aluminum foil etc., accurately.
- It can measure maximum thickness of 25 mm.
- Least count of Digital Micrometer is 0.001 mm.
5.1 PRECAUTIONS
- The Digital Micrometer is a highly heat sensitive instrument and should be stored at room temperature.
- Hold the Digital Micrometer from its Heat Thermally insulating plate whilst handling to prevent heat transfer from hand of user to the Digital Micrometer.
- The object to be measured should also be kept at room temperature since this may also affect the reading.
- Do not over tighten the Digital Micrometer as this can lead to deformation of both the anvil and spindle and the object being measured.
- Store the Digital Micrometer in its appropriate storage compartment to prevent any wearing of the anvil and/or spindle.
- Store the Digital Micrometer with the anvil and spindle not in contact with each other so that any temperature fluctuations will not stress the instrument.
- Measure the desired object three times to ensure that the reading obtained is accurate.
- Do not operate abruptly, drop or strike the Digital Micrometer.
- Digital Micrometer uses 1.55V button cell battery.
- Do not substitute used battery.
- Discard the used battery.
5.2 OPERATION
- Turn the thimble to gently bring the spindle in to contact with the anvil. Apply the rated measuring force using the ratchet stop/friction thimble.
- Press the zero buttons to zero setting display 0.000 on the screen.
- Rotate the thimble anticlockwise allowing sufficient space between the anvil and the spindle to accommodate the object to be measured.
- Place the object to be measured between the anvil and the spindle.
- Spin the ratchet clockwise until the spindle just meets the object.
- Keep spinning the ratchet gently until 3 clicks are heard.
- Check that both the anvil and the spindle are touching the object evenly.
- Set the Spindle clamp while the micrometer is still on the object.
- Note the reading.
- Remove the object from the micrometer.
- Reset the Spindle clamp to the unlock position.
- Keep the instrument in box after use.
- Get calibrated Digital Micrometer by approved outside agency on yearly basis and keep certificate in record.
5.3 CLEANING PROCEDURE
- Rub off any visible dirt, dust and grit with a lint free cloth.
- Wipe off the measuring faces of both the anvil and spindle with lint-free paper set the start (zero) point before measuring.
6.0 ABBREVIATIONS:
Abbreviation | Expanded form |
mm | Millimeter |
7.0 ANNEXURES:
Nil
8.0 SOP REFERENCES
- Manual of Digital Micrometer
END OF THE SOP