SOP for Operation and Calibration of Total Organic Carbon analyzer. (TOC)

SOP for Operation and Calibration of Total Organic Carbon analyzer in Pharmaceutical Industry

  • SOP covers below points :
    • Operation of TOC Analyzer
    • Calibration of TOC Analyzer
    • System Suitability
    • Linearity
    • Safety And Precautions
1.0 OBJECTIVE :
  • To lay down a procedure for operation and calibration of Total Organic Carbon analyzer.
2.0 SCOPE :
  • This SOP is applicable for operation and calibration of Total Organic Carbon analyzer in Quality Control Department.
3.0 RESPONSIBILITY :
  • The QC Staff shall be responsible for:
    • Following the procedure for operation & calibration of Total Organic Carbon analyzer.
  • The Manager-QC shall be responsible for:
    • Ensuring compliance of the SOP
4.0 ACCOUNTABILITY:
  • Quality Control Head 
5.0 PROCEDURE :

5.1 Operation of TOC Analyzer:

  • Open the zero-air cylinder. Switch on the Total organic carbon analyzer. The ‘Initial display’ is displayed.
  • After ready position comes, place the water sample which is to be analyzed.
  • Press ‘Measure sample’ key, the ’Measurement Parameters’ screen is displayed, enter the sample name and press the F5 (NPOC) KEY.
  • Select the calibration curve & run the sample with following parameters:
No of inj.2
Max. of inj.3
SD0.2
CV2.0 %
Dil. Factor1
Acid ratio1.5 %
Sparge time1.0

  • Press the ‘Next’ key & press ‘Start’ key after start key starts blinking.
  • After completion of test, Press F6 (Exit measurement) key & take the print of result.
  • To switch off the instrument press F1 (Standby option) key & then F5 (Power off) key & than press F6 (Execute) key and again F6 (yes) key.

5.2 Calibration of TOC Analyzer:

  • Reagents / Equipments Required:
    • Reagent water for calibration (TOC level less than 100 ppb)
    • Sucrose
    • 1,4-Benzoquinone
  • Frequency:  Quarterly  ± 5 days
  • Preparation of 500 ppb Carbon Solution prepared with Sucrose :
    • Weigh 0.1190 gm of Sucrose, dissolve and dilute to 1000 ml with Reagent Water.
    • Take 1 ml of above solution and dilute to 100 ml with Reagent Water.
  • Preparation of 500 ppb Carbon Solution prepared with 1,4-Benzoquinone :
    • Weigh 0.075 gm of 1,4-Benzoquinone, dissolve and dilute to 1000 ml with Reagent Water.
    • Take 1 ml of above solution and dilute to 100 ml with Reagent Water.
  • Press the ‘Cal’ key for starting the calibration. Press F2 (TC) key.
  • Select the number where Calibration curve is to be saved.
  • Insert the two points with arrow key, take following parameters;
Point 1Point 2
Std. Conc.0.0 ppb500.0 ppb
Inj.22
Max of Inj.33
SD0.20.2
CV2.0 %2.0 %

  • Press the ‘Next’ key & place the Reagent Water & start the measurement. After completion 0.0 ppb concentration, place 500 ppb carbon Solution & press F6 (Go to next Point) key.
  • After completion of calibration press F6 (Data report). Again press F6 (cal. Curves List) key. Select the calibration curve no. where graph is saved. Press F6 (edit) & then F4 (curve shit status) & pass the curve to origin by pressing F4 key.
  • Press F6 (OK) key & press F1 (Return) key.

5.3 System Suitability:

  • Place the 500 ppb of Carbon solution prepared with 1,4-Benzoquinone.
  • Operate the instrument as per operation procedure and find out the area in triplicate.
  • Calculation:
    • Area of 1,4-Benzoquinone – Area of Water / Area of Sucrose – Area of water X 100
  • Acceptance Criteria: Between 85.0% and 115.0%.

5.4 Linearity:

  • Prepare the solution of different concentrations with sucrose.
  • Solution Preparation
    • Stock Solution: Weigh 0.1190 gm of Sucrose, dissolve and dilute to 1000 with reagent water.
    • Dilute 1 ml of stock solution to 200 ml with reagent water. (250 ppb carbon solution)
    • Dilute 1 ml of stock solution to 100 ml with reagent water. (500 ppb carbon solution)
    • Dilute 1.5 ml of stock solution to 100 ml with reagent water. (750 ppb carbon solution)
    • Dilute 1 ml of stock solution to 50 ml with reagent water. (1000 ppb carbon solution).
  • Operate the instrument as per operating procedure and find out area of each solution in triplicate.
  • Calculate % relative standard deviation and relative standard deviation of area of each solution should not be more than 2.0%.
  • Plot a graph of standard concentration against peak area.
  • Calculate the linearity correlation co-efficient.
  • Acceptance criteria: Correlation coefficient should be not less than 0.99.

5.5 Safety And Precautions:

  • Always open zero air gas cylinder before switching on the instrument.
  • The air flow must be 150ml/min & pressure should be 200kpa.
  • Check the water level of humidifier.
 6.0 ABBREVIATIONS:
AbbreviationExpanded form
SOPStandard Operating Procedure
No.Number
QCQuality Control
QAQuality Assurance
DeptDepartment
No.Number
MlMilliliter
MinMinutes
kpakilopascal
TOCTotal Organic Carbon
ppbPart per billion
gmGram
SDStandard deviation

7.0 ANNEXURES:
  Annex. No.Title
01Calibration Protocol for TOC Analyzer
02Usage log for TOC Analyzer

8.0 SOP REFERENCES
  • In-House

END OF THE SOP


ANNEXURES :

Annex. No01 Calibration Protocol for TOC Analyzer


Annex. No02 Usage log for TOC Analyzer